Chamber
size & other specs | Availability | Cost | |
20" x 20" x 38" (New, Recess) | 2 weeks | $33,000 | Click
for details |
16" x 16"
x 28" (New) Electric Control, Steam generator, printer | 4
weeks | $53,497 | Click
for details |
24" x 36" x 48" (New) | 2
weeks | $86,125 | Click
for details |
24" x 36"
x 60" (New) Double Door, Micro Processor Control | In
stock | $65,000 | Call
for details 1-800-748-1259 |
36" x 36" x 60" (New) Single Door, Micro Processor
Control, Steam Generator Also includes ... Steam Generator, Load Car &
Transfare Carriage. | 3 weeks | $59,700 | Call
for details 1-800-748-1259 |
Need
a different size? Click here --
| SPECIFICATIONSELECTRICALAll
operator cycle designations are reduced to 24 volt AC operation. Normal electrical
functions are operated at 110 volt AC. The 24 volt AC operation reduces operator
exposure to electrical power. Most control components shall be "plug-in"
style which enhances service if it should be required. |
Sterilize (Cycle) and Dry Timers shall
be automatic reset type which allows the operator to visibly acknowledge the preset
time as well as the elapsed time.
A main
wiring terminal strip allows easy checking of electrical components if required.
All electrical connectors and wiring shall be
numbered and color coded for ease of service.
Cycle
sequence shall be easily monitored through the use of rectangular indicator lamps
showing each phase of cycle operation.
A
"DOOR LOCKED" light shall illuminate when the chamber door is properly closed.
**A "LOW STEAM PRESSURE" light will illuminate
and the cycle cannot be initiated if a low steam pressure situation exists.
All operation functions shall be initiated through
the use of various solenoid valves which shall fully control the cycle of the
sterilizer.
Cycle selections shall be
designated by the operator by a single press of a button, and the cycle selector
shall illuminate to allow other personnel to realize a cycle has been selected
and started.
If electrical AC power is
unavailable, steam will not enter the chamber. If electrical power is interrupted,
the autoclave shall automatically vent all pressure from the chamber, allowing
the chamber door to be opened. Simultaneously, the unit shall reset all cycle
functions. When power is restored, the cycle selection must again be designated
by the operator.
The entire electrical
circuit shall be protected by a manual reset type circuit breaker, rated for the
maximum acceptable amperage for the electrical components used.
All
electrical components shall be Underwriters Laboratories approved and/or listed,
as necessary.
Temperature control and
documentation shall be assured with the use of a United Electric Model 650, 61'
inkless recorder, mounted in its own cabinet box and secured to the control panel.
PLUMBING
All
piping shall be rigid brass with threaded connections, with the exception of pressure
sensing piping, which shall be 1/4 " O.D. copper with flair or compression connections.
All solenoid operated valves shall be of
one manufacturer, and rated for a minimum of 50 psi operation.
All
plumbing components shall be installed with brass union connectors to allow ease
of service or replacement, if required.
*
* Optional
All manual supply valves shall
be "ball" type, actuated with a manual handle.
Pressure
gauges for supply steam pressure, jacket pressure, and chamber pressure shall
be mounted in such a way that the operator can easily monitor variations.
The vessel shall be protected against high pressure
with the use of a safety valve rated at the maximum operating pressure of the
vessel as stated by the original manufacturer, and shall not exceed A.S.M.E. standards.
VESSEL
The
existing vessel shall be hydrostatically tested at 1.5 times its rated working
pressure to assure long life of the vessel.
The
interior shall be cleaned with the use of a fine glass bead, applied under pressure,
to remove existing discoloration normally found on the inside chamber walls.
The exterior shell shall be painted with high
heat temperature, rust retardant paint.
The
exterior shell shall then be covered with 2' thick, foil backed insulation to
reduce ambient heat normally released from the vessel.
The
door for the vessel shall be refinished and/or rebuilt, as needed.
CONTROL
PANEL
The main control panel shall
be fabricated with .125" thick aluminum. All designation lettering shall be engraved
into the panel and paint filled, for long life and durability.
The
control panel shall be mounted to a fabricated stainless steel primary frame on
top of chamber, and shall tilt down at the operator for ease of visual operation
and a pleasant cosmetic appearance.
The
main control panel shall be hinged for easy access for maintenance and calibration
purposes.
GENERAL
All
stainless steel and painted surfaces shall be refinished as needed for a "like
new" appearance (not applicable to NEW sterilizers).
Complete
calibration and testing shall be performed prior to shipment.
Upon
completion of installation, Alfa shall provide complete orientation and demonstration
of the sterilizer.
Installation is included
only when so quoted. Any structural or cosmetic repairs to existing walls, floors,
ceilings, etc. shall be the responsibility of the customer.