Hirayama
HG Series
HG
With Innovative Feature
Just Press A Button To Automatically Open The
Lid, Or Close And Lock It.
Automatic lid opening button replaces a foot pedal. Automatic lid closing button closes and locks lid conveniently and safely. New body design with a 50- or 80-liter chamber of larger diameter and shorter height. Safety interlock senses pressure and temperature and prevents lid from opening when the chamber is pressurized, temperature is too high, and also during a cycle. Programmable auto-start for running a cycle up to 99 hours later. Efficient 15-liter steam condenser eliminates steam exhaust into the lab. Over-flow lamp indicates that the drain bottle is full and ready to be emptied. Built-in cooling fan for faster post-sterilization cooling and shorter completion time. Selectable rate of steam venting after sterilization. Programmable time for air removal before sterilization. Large control and display panel improves functionality and shows the progress of the sterilization cycle.
|
Three different types of stainless steel containers of different heights (large, medium and small) are available for handling different loads. Stacking shorter containers can increase load capacity (two medium-sized and three small-sized containers can fit in a given chamber). |
Three different types of stainless steel containers of different heights (large, medium and small) are available for handling different loads. Stacking shorter containers can increase load capacity (two medium-sized and three small-sized containers can fit in a given chamber). |
Models HG-50 and HG-80 Specifications |
MODEL | HG-50 | HG-80 |
Outside Dimensions | (19.7W
x 27.2D x 35.0H) in (455W x 691D x 890H) mm |
(19.7W
x 27.2D x 40.6H) in (455W x 691D x 1030H) mm |
Chamber Dimensions | (14.3D
x 18.2H) in, (364D x 482H) mm |
(14.3D
x 28.7H) in, (364D x 730H) mm |
Chamber Volume | 50L working, 60.0L actual volume | 76L working, 85.1L actual volume |
Temperature Range | 105°C to 135°C | |
Sterilization Timer | 1 to 300 minutes | |
Media
Warming/ Holding Temperature |
Programmable at 45°C to 60°C to keep media warm up to 24 hours | |
Auto-Start Timer | Initiate a cycle 1 to 99 hours later | |
Cooling Fan | Switch to turn fan on or off | |
Air Removal Timer | Programmable, 3 to 9 minutes after temperature has reached 97°C | |
Steam Venting | Selectable from: No venting and three other venting rates | |
Temperature for Release of Lid Interlock | Programmable, 60°C to 97°C for greater safety and flexibility | |
Sterilization Program Cycles |
1) Liquid cycle, 2) Agar (liquid with warming) cycle, and 3) Solid cycle | |
Additional Cycle for Agar Dissolution | Cycle for dissolving and keeping agar warm (no sterilization) | |
Agar
Dissolution Protocol |
Initially at 60°C-100°C for 1-60 min., then at 45°C-60°C for up to 24 hrs. | |
Safety Features | Circuit breaker with GFCI, low-water power cut-off, over-temp power cutoff, over-pressure power cut-off, safety valve, sensor for improper lid closing | |
Steam Condenser | 15-liter water tank | |
Drain Water Bottle | 2 liters | |
Power Requirements | 120V or 220V, 50/60 Hz, single phase | 220V, 50/60 Hz, single phase |
Power Consumption | 2.0kW, optional 3.0kW for 220V only | 3.7kW |
Shipping Weight | 154 lb (70 kg) | 181 lb (82 kg) |