MATERIAL SAFETY DATA SHEET 077-0033-00 REVISION G
SECTION 1- PRODUCT IDETIFICATION
Product Name: SPEEDCLEAN Date of Preparation: May 12, 1997
Date of Review: October 26, 2000
Manufactured for: Midmark Corp.
Versailles , Ohio 45380
937-526-3662 /Fax 937-526-4604
SECTION 2- HAZARDOUS INGREDIENTS
Ingredients: CAS# ACGIH TLV %(w/w)
2-butoxyethanol 111-76-2 25ppm 2-5
SECTION 3- PHYSICAL DATA
Specific Gravity: 1.03
Boiling Point: 212 degrees F.
Vapor Pressure: 17.5 mm Hg.
Vapor Density: NA
Evaporation Rate: same as water
Solubility in Water: complete
Appearance: Yellow liquid
Odor: Mild Solvent
SECTION 4- FIRE AND EXPLOSION HAZARD DATA
Flash Point (Deg.F.): None through boiling
Lower Explosive limit: None
Upper Explosive limit None
Extinguishing Media: water,foam,carbon dioxide,dry
chemical as for surrounding fires
Special Fire Fighting Procedures: None
Unusual Fire/Explosion Hazards: None
SECTION 5- REACTIVITY DATA
Chemical Stability: Stable
Conditions to avoid: Extremely High Temperatures
Incompatible Materials: Strong oxidizers, acids
Harzadous Decomposition Products: Carbon dioxide,carbon monoxide
Hazardous Polymerization: will not occur
SECTION 6- HEALTH HAZARD DATA
Effects of Overexposure:
Eyes - Contact can cause irritation, redness.
Skin - Prolonged or repeated contact can cause irritation.
2- butoxyethanol is aborted through the skin and
chronic overexposure may cause blood disorders,
liver or kidney injury.
Breath - Mist may irritate nasal and respiratory passages.
Acute overexposure to vapor may cause central
nervous depression. Chronic overexposure to
vapor may cause blood disorders, liver, or kidney injury.
Eyes - Flush them with water for 15 minutes.
Get medical help if irritation persists.
Skin - Rinse well with water if contact occurs.
Respiratory System - Remove to fresh air. If necessary, give oxygen, artificial
respiration.
Ingestion - Give large quantities of water .Get medical assistance.
Carcinogenicity - Not listed as carcinogenic by LARC, NTP, OSHA, ACGIH.
SECTION 7- PRECAUTIONS FOR SAFE HANDLING AND USE
Steps to be taken in case of spill: -Absorb on solid absorbent or wash down with water.
Waste Disposal Method: -Dispose in accordance with all local, state, and federal regulations.
Precautions to be taken in handling and Storing: -Store at moderate temperatures keep container closed when not in use.
Other precautions -Keep out of reach of children.
Return empty drums to licensed
Reconditioning service.
SECTION 8- SPECIAL PROTECTION INFORMATION
Respiratory Protection: -use NIOSH approved respirator for product
Mist and vapors.
Ventilation: - Recommended
Protective Gloves: - Required
Eye Protection: -Wear safety glasses or face shield.
Other protective clothing and equipment: -Eye wash station
SECTION 9 – HMIS RATINGS
Health: 1 Fire:0 Reactivity:0
SECTION 10 – L313 LISTED MATERIALS
As per the requirements of 40 CFR 372.45, the following components of this product are
Listed in this section 313 of Title 3 of the Superfund Amendment and Reauthorization Act of 1986.
Material CAS# Maximum% by weight
2-butoxyethanol 111-76-2 3
This information is based on data available to us and is accurate and reliable to the best of our knowledge at the time of printing. However, no warranty is expressed or implied regarding the accuracy of completeness of the information contained herein. Final determination of the suitability of this material for the use contemplated is the sole responsibility of the user. Buyer assumes all risk and liabilities. Buyers accept and uses this material on these conditions.
Instructions on how to use the Speed Clean:
1. Drain and refill the reservoir with clean distilled or demineraized water.
2. Add one (1) ounce of SPEED CLEAN to a cool chamber
3. Run one (1) wrapped/pouched cycle or one (1) cycle at 270f (132c) for six (6) minutes. Instruments should not be sterilized while cleaning sterilizer.
4. After cleaning, drain and refill the reservoir with distilled or demineraized water. Run one (1) unwrapped cycle or one (1) cycle at 270f (132c) for three (3) minutes.
5. Drain reservoir and allow sterilizer to cool to room temperature. Remove the trays, tray rack, and tray plate. (See the operators manual). See the operator's manual for gasket care and cleaning. Wipe out the inside of the chamber being careful not to damage the heater element or the temperature and level sensors. Wipe off the tray components and reinstall in the chamber. Make sure the tray rack is to the back of the chamber.
6. Refill reservoir with clean distilled or demineraized water.
7. The sterilizer is now ready to for use.
Re ordering:
Midmark Speed Clean autoclave cleaner one case
For assistance call 1- 800-801-9934 and ask for
Andrea, Renata, Ed, Shlomo, or Chuck
Contact Us
Find an autoclave or sterilizer that will sterilize all instruments. COX dry heat sterilizer, Pelton and Crane OCM, OCR; Distributors of Sterilizers & Autoclaves. |