Error Code |
Symptoms |
Reason |
How to Fix |
Part to Replace |
Part Number |
E01 |
The cassette temperature failed to reach 95°C in 3 minutes. |
Steam generator does not heat up. No power to steam generator. Blown thermal fuse. |
Check for a large cassette leak or a large load. Replace the thermal fuse if necessary. |
Thermal fuse
|
01-104461S
|
E02 |
The cassette temperature failed to increase from 95°C to 100°C within 1 minute 20 seconds. |
Faulty cassette seal, miscalibration, damaged cassette, improperly adjusted or failed microswitch allowing unit to function when cassette is not fully inserted. |
Replace the cassette seal, calibrate the unit, or repair/replace the cassette as needed. |
Cassette seal
Microswitch
Cassette
| 01-100028S
01-100571S
01-100008A
|
E03 |
The cassette has failed to pressurize and achieve a temperature of 110°C within 70 seconds of pressurization. |
Faulty or worn cassette seal, faulty or dirty solenoid valve, damaged cassette, or faulty exhaust tube, leaky check valve or pressure relief valve. |
Replace the cassette seal, clean the solenoid valve, repair/replace the cassette as needed, or repair/replace the exhaust tube, check valve, or pressure relief valve as needed. |
Cassette seal
Solenoid valve
Cassette
Exhaust tube
Check valve
Pressure Relief valve
|
01-100028S
01-100557S
01-100008A
MZZA100016
01-100574S
01-106787S
|
E04 |
The cassette has failed to achieve sterilization conditions within 10 minutes of the chamber first reaching 102°C. |
Damaged or worn cassette seal, faulty solenoid valve, improperly installed copper tubing, cassette thermocouple misalignment or damage, faulty steam generator pressure relief valve, or faulty check valve on. |
Replace the cassette seal, clean the solenoid valve, reinstall the copper tubing correctly, align or repair the cassette thermocouple, repair/replace the steam generator pressure relief valve, or repair/replace the check valve on the steam generator as needed. |
Cassette Seal
Solenoid Valve
Copper tubing
Cassette thermocouple
Pressure relief valve
Check valve
|
01-100028S
01-100557S
01-109128S
SCT030
01-106787S
01-100574S |
E05 |
The boiler went dry prematurely. |
Bad pump or boiler. Check pump using Pump Tester Bottle Part #SCK024-384.
|
Check the pump and replace it if necessary.
|
Pump
|
01-100568S
|
E06 |
The boiler temperature is more than 5°C higher than the chamber while sterilizing. |
Check for kinks in the exhaust tubing. Check for a clogged duct in the rear of the cassette tray. Blow out with Compressed air. Check solenoid for proper operation. |
Check the exhaust tubing, clean the duct in the rear of the cassette tray, and/or check the solenoid valve for proper operation. |
Exhaust tubing
Solenoid Valve
|
MZZA100016
01-100557S
|
E07 |
The cassette has failed to pressurize. |
Faulty or dirty solenoid valve, damaged cassette, or faulty exhaust tube. |
Clean the solenoid valve, repair/replace the cassette as needed, or repair/replace the exhaust tube as needed. |
Solenoid valve
Cassette
Exhaust tube |
01-100557S
01-100008A
MZZA100016
|
E08 |
The cassette has failed to sterilize. |
Damaged or worn cassette seal, faulty solenoid valve, improperly installed copper tubing, cassette thermocouple misalignment or damage, faulty steam generator pressure relief valve, or faulty check valve on. |
Replace the cassette seal, clean the solenoid valve, reinstall the copper tubing correctly, align or repair the cassette thermocouple, repair/replace the steam generator pressure relief valve, or repair/replace the check valve on the steam generator as needed. |
Cassette seal
Solenoid valve
Copper tubing
Cassette thermocouple
Pressure relief valve
Check valve
|
01-100028S
01-100557S
01-109128S
SCT030
01-106787S
01-100574S
|