Top-Loading Autoclave With Fully Automatic Drying
The HVP-50 autoclave, with a 50-L chamber, is a self-contained, portable floor model mounted on caster wheels. It occupies minimum floor space, operates at 120V, and can fit conveniently into many locations in the laboratory. Its many uses include the sterilization and drying of glassware and pipette tips, and the sterilization of wastes and liquids such as agar and media. The HVP-50 model can accommodate a 10-L fermenter, a 15-L carboy, a 19 x 24 biohazard bag, six 1-L flasks, twenty-one 500-mL bottles, or one hundred eighty 20-mm test tubes. Space-saving design with lid opening upward. Fully automatic drying
cycle for solid loads uses
heat and vacuum for faster and more efficient drying. or when the temperature exceeds 97°C. Vacuum pump removes air from the chamber, when sterilizing solids such as glassware and pipette tips, thereby eliminating
cold spots, and ensuring faster and more efficient sterilziation.
Liquid
cycle for media and agar. Cycle
for wastes and glassware without drying.
|
Three different types of stainless steel containers of different heights (large, medium and small) are available for handling different loads. Stacking shorter containers can increase load capacity (two medium-sized and three small-sized containers can fit in a given chamber). |
HVP-50 Specifications |
TEMPERATURE | |
Operating Range | 105ºC to 135ºC |
Display Range | 5ºC to 140ºC |
GAUGE PRESSURE | |
Operating Range (gauge) | 3 to 28.4 psi (0.2 to 2.0 kg/cm² ) |
TIMER | |
Sterilization | 1 to 250 minutes |
Drying | 1 to 90 minutes |
DIMENSIONS | |
Overall (W x D x H) | (20.1 x 24.8 x 40.9) in (51 x 63 x 104) cm |
Chamber (Diameter x Height) |
50
liter 11.8" x 28" (30 cm x 71 cm) |
SAFETY DEVICES | Temperature- and pressure-sensitive lid interlock, over-pressure power cutoff, over-temperature power cutoff, low-water power cutoff, mechanical pressure relief valve, lid-closure detector, over-current detector and circuit breaker with ground fault circuit interrupter. |
PROGRAMS | |
1. For solids with drying | Heating and air removal from chamber by vacuum pump (pre-vac) -> Sterilization -> Water and steam exhaust (fast) -> Drying by heat and vacuum -> Completion |
2.
For wastes/solids without drying |
Heating -> Pre-vac -> Sterilization -> Steam exhaust (fast) -> Completion |
3. For liquids/agar | Heating -> Sterilization -> Auto steam exhaust (programmable rates, or ambient cooling w/o steam exhaust) -> Completion |
4.
Drying without sterilization |
Drying with heat and vacuum (time programmable) -> Completion(for additional drying after Program #1, if desired) |
ACCESSORIES | |
Stainless steel | Mesh wire baskets, buckets (pails), and test tube rack |
POWER | |
Source | 120V, 50/60 Hz, single phase (220V/240V available on request) |
Consumption | 2000 watts (2000 VA) |
SHIPPING WEIGHT | 291 lb (133 kg) |
For assistance call 1- 800-801-9934 and ask for
Andrea, Renata, Ed, Shlomo, or Chuck
Contact Us
Find an autoclave or sterilizer that will sterilize all instruments. COX dry heat sterilizer, Pelton and Crane OCM, OCR; Distributors of Sterilizers & Autoclaves. |